FDA Recalls St. Jude’s Riata Lead Based on Risk of Death

St. Jude Medical Inc.’s Riata defibrillator leads, which the company stopped selling last year, were recalled by the U.S. Food and Drug Administration because of their potential to injure…

St. Jude Medical Inc.’s Riata defibrillator leads, which the company stopped selling last year, were recalled by the U.S. Food and Drug Administration because of their potential to injure or kill patients.

The devices remain implanted in an estimated 79,000 U.S. patients. The company voluntarily sent a letter to doctors on November 28 informing them that the wires used to connect the devices to the heart have a higher failure rate than was previously known.

According to the company, the FDA issued the Class 1 recall, its most serious designation, because of its potential risk of serious injury or patient death.

The company acknowledged that the wires inside some leads can penetrate the insulation, compromising the device’s integrity. The devices may inappropriately shock some patients and fail to deliver necessary therapy to others.

This potential for device failure, particular for defibrillators or for folks dependant on pacemakers, is worrisome.

Patients who need frequent shocks from their defibrillator to restore a normal heart rhythm and those who are dependant on pacemakers may be at greater risk, the company said. Doctors may design individual treatment plans for those with the damaged wires based on their risk level and options should be discussed with patients, the company said.

FDA notified healthcare professionals of a Class I Recall of the St. Jude Medical  Riata and Riata ST Silicone Endocardial Defibrillation Leads. The silicone insulation covering these defibrillation leads is at risk of premature abrasion. When abrasion occurs, the conductors inside the leads can come out (externalized) of the insulation. Leads with externalized conductors may develop electrical dysfunction and not work as intended. In the event the device does not work as intended, should a life-threatening heart rhythm occur, pacing or defibrillation therapy may not be delivered as intended. This may result in serious adverse events, including death.

The recall includes the following model numbers:

Riata (8F) Silicone Endocardial Defibrillation Leads
     Models: 1560, 1561, 1562, 1570, 1571, 1572, 1580, 1581, 1582, 1590, 1591, 1592
     Riata ST (7Fr) Silicone Endocardial Defibrillation Leads
     Models: 7000, 7001, 7002, 7010, 7011, 7040, 7041, 7042

BACKGROUND: The leads connect an implantable cardioverter defibrillator (lCD) or cardiac resynchronization therapy defibrillator (CRT-D) to cardiac tissue in order to monitor and regulate a patient’s heart rate by providing pacing and delivery of high voltage therapy for ventricular arrhythmias.

Shezad Malik MD JD

Shezad Malik MD JD

Shezad Malik is an Internal Medicine and Cardiology specialist, a Texas Medical Doctor (retired) and Defective Medical Device and Dangerous Drug Attorney.

All articles

More in Medical Devices & Implants

See all

More from Shezad Malik MD JD

See all
Ozempic and Wegovy, Here’s the Skinny

Ozempic and Wegovy, Here’s the Skinny

/
Benzene Work Exposure leads to Blood Cancer

Benzene Work Exposure leads to Blood Cancer

/